Pendulum measures periodic limb movements in sleep from a smartwatch worn at the ankle, and tracks the rhythm between them rather than their number. It is it measures, it does not interpret — a real measurement to take to a physician, not a diagnosis and not a medical device — see what this is, and what it is not.
These are real screenshots. They were captured with adb exec-out screencap from the application
running on emulators: an Android 14 phone profile, and a small round Wear OS 5 target for the watch. The interface is in English; the working documents under docs/fr/ are not, and that is
deliberate — they are the record of how the decisions were made, not user-facing text.
Every number visible on them was produced by the application's own preview
dataset flowing through the real Compose code, and every chart was drawn by the same DrawScope
extensions the production build uses.
What they are not: a recorded night. No accelerometer data has ever been captured at an ankle by this software, so the plotted signal is the synthetic preview data, and the indices are the values that dataset produces. The screens are real; the night is not.
Hand-authored SVG mock-ups of the same screens are kept in images/ for the three cases the
emulator run did not reach — the first-run warning, the evening card, and the trend below three nights.
They are drawings, not captures, and they are labelled as such where they appear.
Three defects were invisible until the application actually rendered, and all three are the kind a drawing cannot expose because the person drawing it lays the text out by hand:
- Axis labels overlapped into an unreadable block on all three charts.
axisTextSizewas stored pre-multiplied by screen density and then consumed as.sp, which applies density a second time: a requested 11 sp was drawn at 83 px on a 2.75-density screen. Fixed inChartTokens. - Text was clipped by the curve of the round watch face. A fixed 16 dp padding is fine on the
square previews the toolkit renders and wrong on a circle. The margin now derives from the screen
shape; it took two attempts, because the inscribed-square inset is only correct for static content
and this column scrolls. See
RecordScreen. - The watch and phone declared different application identifiers. The Wearable Data Layer only exchanges between applications sharing an identifier, so the two halves would never have seen each other — and the failure mode is silence, not an error.
| Faithful | Source |
|---|---|
| French strings | fr/UX.md, verbatim wherever it wrote them |
| Dark palette, exact hex | fr/UX.md §5.1 |
| Spacing scale, 14 dp card radius, 1 dp borders, no shadows | fr/UX.md §5.3 |
| Layer order, marker shapes, axis conventions of the plots | fr/UX.md §4 |
Not faithful, and deliberately so: the typeface. The specification calls for Roboto Flex with tabular figures; these files use a system font stack so that they render the same in a browser, in a Markdown preview and on GitHub, none of which will fetch a font. Line lengths are therefore approximate and body text is one to two points smaller than the 15 sp the specification asks for, which is what let four full paragraphs fit on the warning screen at a legible size.
Three strings on these screens are not in the specification and were written for the mock-up. They
are marked in the sections below: the evening-record form, the sentence explaining why the result is
hidden on waking, and the two captions under the trend plot. Each corresponds to a decision recorded in
01-overview.md §4 that never reached fr/UX.md, which was written
before it.
Step 1 of 5, in a pager that does not swipe. Progression is by button only, and the button is disabled until the text has been scrolled to the bottom — which is why the mock-up shows it mid-scroll, greyed, with the paragraph above it cut by the top edge and one more paragraph waiting under the fade.
The mechanism matters more than the words. A warning you can dismiss with a thumb-flick has been dismissed by everyone within a week; a warning that costs a scroll and a deliberate tap has at least been scrolled past deliberately. The same four limits are reproduced verbatim in the exported PDF, so a physician reading the report sees exactly what the user agreed to at install time.
The limits are stated as permanent — quatre limites qui ne disparaîtront jamais — rather than as caveats, because three of them are properties of the sensor and its position, not defects to be fixed in a later version. No amount of signal processing gives an ankle accelerometer a respiratory channel.
Two hands: the watch starts the recording, the phone holds the record of the evening. The top card is the one the specification draws — watch battery and free space, the strap marker and which leg, whether the sleep source is active — and it exists only between 20:00 and 04:00. Outside that window it is not collapsed or greyed, it is absent, so that nothing useless is on the screen at 07:00.
The battery line is advice: below 85 % it turns amber and says 98 % recommandé — une nuit consomme 40 à 70 %, and it blocks nothing. The thing that does block is the second card. The watch refuses to
start until the evening record is sealed, append-only, not editable afterwards — the first guard rail
in 01-overview.md §4. Its purpose is narrow and worth stating: it prevents the
context from being adjusted after the result is known. Someone who sees a bad night and then remembers
they had a drink is not being dishonest; they are being human, and the only place that can be made
structurally difficult is in the code, before the number exists.
The fields of this card are written for the mock-up. 01-overview.md names them — dose, context, which
leg, which strap, alone in the bed — but fr/UX.md predates the decision and never wrote the screen.
This is the screen the user sees most mornings, and it is the one most likely to be mis-designed. The movements have been scored; the wrist device's sleep stages have not reached Health Connect yet, and may not for hours — the manufacturer's own documentation describes watch-to-phone transfer as governed by the watch's battery policy, with no guaranteed delay.
So the design constraint is: this must not look like a failure. Nothing on it is red. The title is
neutral and factual, there is no error code, and the flag is an amber masque accéléro chip, which is a
flag and not a diagnosis of the app's health. The body text says outright that this synchronisation
often happens several hours after waking and that the figure will be recomputed automatically with
nothing to do. Réessayer maintenant exists for the impatient; the retry ladder — 30 min, 1 h, 2 h, 4 h,
8 h, give up at T+36 h — runs whether it is pressed or not. If the interface presented a missing
hypnogram as an anomaly, the user would conclude the app was broken every single morning, and would stop
believing the one flag that does mean something.
The second card is the guard rail that hides the aggregate at waking. Revealing it is a deliberate tap and is written to the technical log, and the log goes into the export. That sentence is written for the mock-up; the guard rail is specified, its wording is not.
Read it from the top, because the order is the argument.
The verdict comes before the number. The first line says the variation is indistinguishable from night-to-night variability; the figure follows. Reverse those two and the number is read and the caution is not. The words amélioration, aggravation, efficace exist nowhere in the string resources, and a unit test asserts that on the resource file itself.
The number is a rhythm in seconds, not an hourly count. Night-to-night variability of the mean log inter-movement interval is 3.6 % against 43.2 % for the hourly count (Skeba et al. 2016) — twelve times more stable, and it needs no denominator. The hourly count stays, one line down, in tertiary type, labelled pour le médecin, because that is the language sleep physicians read.
The uncertainty is measured, not decorative. The CI band is a percentile bootstrap over the user's own nights with a fixed seed, so the interval does not move between recompositions. The wider band is the minimum detectable change: below it, a difference is not a difference. Both bands are labelled in text and bordered with different dash patterns, so neither depends on its fill colour to be identified.
The points are not joined. A polyline between nights asserts a continuous trajectory and a causality that do not exist; if a monotone trend is really there, the positions of the points will show it. The excluded night is drawn at its value as a hollow circle with its reason beside it, and counted in nothing — hiding the failed nights would distort the reading of the whole. The X axis is calendar, not ordinal, so the three days without a recording leave visible holes. Those holes are information.
Two departures from 06-interface.md are visible here and both are open questions
rather than oversights, recorded in §4.5 of that document:
- The Y axis does not start at zero. The "axis starts at 0" invariant was written for a per-hour count. A rhythm in seconds has no meaningful zero, and anchoring it at zero would compress every night into the top fifth of the plot. The axis instead spans a fixed window around the data, and the caption says so — pas d'ancrage à zéro : un rythme n'en a pas. The convention still needs to be specified properly; this mock-up shows one plausible answer, not a decided one.
- There is no 15/h reference line, because there is nothing to draw. The 15/h threshold belongs to the hourly count and does not transfer to a rhythm, and the published periodicity threshold (≈ 0.5, Ferri et al. 2016) is on a different scale. Drawing a line here would be inventing a threshold.
For the same reason the headline sentence is the minimum-detectable-change statement rather than one of the five position sentences of §4.4: those five attach to the 15/h boundary, and the equivalent sentence for the fundamental rhythm has not been written yet. The two captions under the plot are mock-up copy.
This screen matters as much as the previous one, and it is the one that would be cut first by anyone optimising for a demo.
Below three eligible nights, Pendulum does not compute a median, does not store one, does not export one, and does not draw the plot at all — not even empty with axes, because an empty axis invites the eye to imagine a curve. This is not a warning shown next to a disabled feature; there is no code path that produces the value. The distinction is the whole point: a warning is advice you can decline.
Three details are load-bearing:
- Dots, not a progress bar. A bar suggests a score that is rising and rewards the user for filling it. Two filled discs and one empty one state a fact about how many nights exist.
- The reason is given in figures. Le seuil de 15 mouvements par heure n'est dépassé que sur environ une nuit sur trois — even in people whose disorder is confirmed. The intended user is technical, and a quantified reason persuades where an instruction irritates.
- The export button stays visible and disabled, with its reason written on the button itself. Never an enabled button that fails, and never a silently missing one.
The link into each night's detail stays live. The refusal is about aggregating, not about looking: the per-night figure exists, it is simply in the one place where it cannot be mistaken for a result.
The header fuses the single-night value with its warning into one block, so the figure cannot be screenshotted without the sentence that qualifies it. It is body-sized and in secondary colour: available to the technical user who needs it to check that the measurement worked, never a headline.
The chart is six layers. Y is amplitude divided by the night's own noise floor, on a log₂ scale, which is what makes the ×8 onset threshold a straight horizontal line and the detector's logic readable at a glance; useful amplitudes span 1.5× to 30×, and on a linear axis the small events disappear. The envelope is decimated min/max per pixel column and never averaged — a 40 ms peak has to survive eight hours being drawn in 300 pixels, and a unit test asserts it. That is why the envelope looks like a picket fence rather than a smooth curve: each vertical stroke is one column's minimum-to-maximum, which is what an accelerometer envelope with a noise floor and short bursts actually looks like at this scale.
Event markers sit on their own 14 dp band under the signal area, never overlaid on the curve: solid ticks for counted movements, hollow ticks for movements in wakefulness, fine crosses for posture exclusions, and a continuous bar underneath for each series of four or more. The signal gap is drawn at its true width — 3.4 s on a 7 h 46 axis is about one pixel — and annotated, rather than widened to look impressive.
The hypnogram sits directly below, same width, same X transform, three lanes: the accelerometric mask,
the Health Connect stages in laboratory order with wake at the top, and a lane of amber hatching where
the two disagree. REM is the only stage with a distinct treatment, hatched rather than merely coloured,
because few periodic movements are expected there. Where no hypnogram exists that middle lane is not
drawn empty; it is replaced by a muted band carrying Hypnogramme indisponible — masque d'immobilité accéléro utilisé.
The quality block lists each check with its measured value, its threshold and its state. A single failed check makes the night ineligible, and the reason is then shown at the top in amber — not red, because a short night or an unworn watch is a situation, not a fault.
One screen. No navigation, no swipe, no list, no complication, and no result of any kind, ever. Pure black, because on OLED black does not light the pixel, and that is battery life rather than style.
Everything on it is a correctness check, read at most twice a night: elapsed time in large tabular figures, sample count, measured rate against the 50 Hz requested, gap counter, bytes written and space left, battery, and the FIFO batching mode actually in force. Stopping asks for a confirmation, because an accidental press at three in the morning costs the whole night.
The screen refreshes every 30 s exactly, and only while the display is on. There is no animation anywhere, for three reasons in order of weight: every recomposition wakes the SoC and over eight hours a discreet 60 fps animation costs more than the 50 Hz acquisition itself; an interface that invites you to look at your ankle invites you to bend over and move the leg being measured, which contaminates the measurement; and there is nothing to watch in real time anyway. The screen is boring on purpose.
The seven principles of 06-interface.md §1 were written to be answerable yes or no
by looking at a screen. Five of them can be checked on these files as they stand.
| # | Constraint | Where to check it |
|---|---|---|
| P1 | No aggregation below three eligible nights | trend-refused.svg — no plot, no median, no category, export disabled with its reason |
| P2 | Every aggregate figure carries its uncertainty in the same line | trend.svg — IC 95 % : 19 – 23 s · 7 nuits éligibles, and the hourly count carries its own |
| P4 | At most three pieces of information above the fold on waking, at most one action | waking-provisional.svg — status, hidden aggregate, one button |
| P5 | No truncated scale, no mean hiding an extremum | night-detail.svg — min/max decimation; and the trend's zero-anchor exception, argued above rather than assumed |
| P6 | Colour is never the sole carrier | Desaturate any of these files: night states stay distinguishable as filled / ringed / hollow, chart lines as solid / dashed / dotted, REM and disagreement by hatching, sleep stages by vertical position |
| P7 | The per-night figure is available but never foregrounded | night-detail.svg — body size, secondary colour, fused with its warning; absent from trend.svg entirely |
P3 — a variation indistinguishable from noise is named as such before it is quantified — is visible in
trend.svg, where the statement precedes the figure, but its real test is a unit test on the strings
file, not a picture.
The one thing a mock-up cannot check is the thing most likely to go wrong: whether these screens survive a font scale of 1.3, a night with no data, a period containing four nights, and a physician who wants the number. That check needs the real thing.